53
e700 User Manual
AC/DC Power Supply
Model PMP90-13-2 01CV0105
Input Voltage 100 - 240 VAC
Input Current 1.06 - 0.45 A
Input Frequency 47 - 63 Hz
Output Voltage 19 VDC
Output Current 4.74 A Maximum
DC Output Plug 2.5 x 5.5 x 11 mm
DC Output Cable Length 6’
AC Power Cord Length 6’
Weight 642 g
EMC performance Meet requirements of IEC60601-1-2
Safety Standards
Meet requirements of
IEC60601-1:2005
Note: Upon disconnecting AC Power supply, Ventilator will
automatically switch to Battery operation without a ecting
ventilator behaviour.
9.4. Battery Operating Time
Battery operation time ranges from 22 hours in normal operating
temperature to 16 hours under extreme low operating temperature.
9.5. Electromagnetic Compatibility
O-Two e700 has been tested and complies with IEC 60601-1-2:2007
requirements.
Electromagnetic Emissions
O-Two e700 is intended for use in the electromagnetic environment
specifi ed below. The user of O-Two e700 should assure that it is not
used in environments outside those specifi ed: