9.6. Oxygen Consumption
For a “D” size cylinder (capacity of 425 liters), pressurized to 2015
PSI and with the e700 set to the Adult default setting (Vt= 500ml,
Rate= 10BPM, 100% Oxygen) the pneumatic operating time is 85
minutes without PEEP (0.2 min/L) and is 39 minutes with maximum
PEEP (0.089 min/L).
Duration of consumption of a cylinder in minutes is calculated
approximately by dividing the liter content by the minute volume
with or without PEEP.
RATED
MAXIMUM
OUTPUT
POWER OF
TRANSMITTER
W
SEPARATION DISTANCE (M)
150kHz to 800
MHz outside ISM
bands d = 3.5/3
*√ P
150kHz to 800
MHz in ISM bands
d = 1.2√ P
80 MHz to 800
MHz
d = 1.2√ P
80 MHz to 800
MHz
d = 1.2√ P
0.01 0.12 0.12 0.12 0.23
0.1 0.4 0.4 0.4 0.7
1 1.16 1.2 1.2 2.3
10 3.8 3.8 3.8 7.3
100 12 12 12 23