7
e700 User Manual
The external power supply and battery are component parts of
the medical electrical equipment system.
The ventilator is considered as a high fl ow device as its
maximum fl ow output at pressure of 40.6 PSI is around 100
L/min. It should only be connected to a pipeline system that
allows for the indicated high fl ow to avoid interfering with the
operation of adjacent equipment.
Always ensure that all components are assembled correctly
and ready for use.
When selecting very small tidal volumes during ventilation of
infants, take into consideration the dead space in the patient
circuit.
2. Intended Use
The e700 is a time-cycled, volume-constant and pressure-controlled
emergency and transport ventilator designed for use in the pre-
hospital, intra-hospital, inter-hospital and transport settings. It is
intended for use with Adult, child, infant patients with a tidal volume
from 50 ml upwards who are in respiratory and/or cardiac arrest or
respiratory distress and who require the ventilatory support.