EasyManua.ls Logo

OEC COMPACT 7600 Series - Fluoro Mode Logic (B104 Rev B), p. 2, 3;97

Default Icon
54 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
CONTROL PANEL
U2
FLUORO EXPOSURE TIME
DISPLAY
ST3
Xray Indicator
H3 H4 H5 H6
10
Min
1
Min
10
Sec
1
Sec
6
7
8
9
10
11
12
13
16V
-XRAY
16
OPTO
ISOLATOR
U8
HAND SWITCH
FOOT SWITCH
-HAND
-FUSS
OPTO
ISOLATOR
U9
3
6
6
3
-HANDSW
-FOOTSW
ST20
ST10
B106 BOARD
2
1
2
1
D15
CONTROL
LOGIC
PAL U2
PAL U3
U11, U12
RELAY K1
2 second
TIMER U14
OPTO
ISOLATOR
U6
OPTO
ISOLATOR
U7
OPTO
ISOLATOR
U2
OPTO
ISOLATOR
U13
-HANDSW
-FOOTSW
D14
ST6
5
2
-FLUOREQ*
-MEMFLUREQ
-GERR
D53
MONOTOR
XRAY LAMP
LOGIC
RELAY K1
INVERTER U8
CONTROL
LOGIC
PAL U1
B100 BOARD
B104 BOARD REV B
-FLUOREQ*
-MEMFLUREQ
-XRY
-STRA
-SPT
-SPR
-TRIGIN
-TRIGOUT
2
16
10
8
26
6
21
44
2
6
16
3
U11-10
U3-4
ST5 ST1
11
10
11
10
U8-8
24VAC
K1-1
From B111MR board X4-2
(see Monitor Rotation diagram)
To/from image processor
(see Image System diagram)
MONITOR ASSEMBLY
XRAY TUBE HEAD
ASSEMBLY
FILTER BOARD
XRAY LAMP
B103 BOARD
ST4
4
4
X1
X4
1
24VAC
X3
3
4
10ER
TIMB
100ER
1000ER
1ER
TIMA
TIMD
TIMC
6
7
8
9
10
11
12
13
ST2
B101 BOARD
FLUORO TIME
DISPLAY LOGIC
PAL U17
PAL U18
U4, U10
U11, U12
U13, U14
Jumper BR2
IN = 4 sec fluoro beep
OUT = no flouor beep
Jumper BR1
Always out
Jumper SCH1
1
2
3
-XRAY
To
kV control logic
(see kV Generation
& Regulation diagram)
B64 BOARD
To
mA control logic
(see mA Generation
& Regulation diagram)
B143 BOARD
K1-14
ST1
16
18C
-XRAY
-XRAY
X1
18C
X1
11A
ST1
18C
ST4
-XRAY
-GERR
-XRAY
34
1
1
ST1
17
16V
LED 4
S7
Auto
Fluoro
Mode
-KFLUORO
U12-4
U12-2
-KFLUORO
17
ST2
D20
U12-15
U23-1
U10-6
5 Minute Warning Indicator
16V
-5MINLED
36
S28
35-5MINSW
36
35
29A
29C
U10-4
U18-23
ST1
29C
29A
-LEDALARM
-CLRBUZ
See
Interface
Diagram
-GERR
(Generator Error)
(See page 1 for REV A board)
Page 2 of 2
For Reference Only
FLUORO MODE CONTROL LOGIC
3/7/97
Fluoro_C.ds4
(Using the B104 Rev B board)
COMPACT 7600
F1
1A
Power for board
ST1
23C
-BUZZER
Mother
Board
ST9
16VDC
D43
Xray-On (Green)
U11-2
1
TRIGIN = enables the image processor.
TRIGOUT = from the image processor stays active, enabling exposure, for
640 msec after the xray switch is released. This ensures sufficient time to for last
image hold. The timer U14 on the B100 board terminates exposure after 2 seconds
if the TRIGOUT does not terminate after the 640 msec time period.
SEE PAGE 1 FOR SYSTEMS WITH B104 Rev A
NOTES:
Used and Refurbished C-Arms are Available from www.SharpMedical.com - Call us at 800-969-9800
An Independent C-Arm Service Provider. This PDF provided for research / historical purposes only.

Table of Contents

Related product manuals