What to do if OHAUS Adventurer Scales show Err 3.0?
- JJohn MasseySep 7, 2025
Err 3.0 signifies a calibration error, specifically a wrong mass. This is due to an incorrect or missing calibration mass during the calibration procedure. You should recalibrate correctly.