10 Technical specifications
57
Type of gas
Measurement
range
(ppm)
XP
Version
IS Version
Temperature
range °C
% RH
(at 20°C)
Accuracy
(ppm)
Lyfe
(months)
Reponse
time
T
50
/ T
90
(s)
Storage
conditio
ns
Warm-
up time
max (h)
AsH
3
Arsine
1.00
-20 to +40
20 - 90
+/- 0.05
18
30/120
(1)
1
CH
2
O
Formaldehy de
50.0
-20 to +50
15 - 90
+/- 1.5
36
50/240
(1)
36
Cl
2
Chlorine
10.0
-20 to +40
10 - 90
+/- 0.4
24
10/60
(1)
1
ClO
2
Chlorine Dioxide
3.00
-20 to +40
10 - 90
+/- 0.3
24
20/120
(1)
1
CO
Carbon monoxide
100
300
1,000
-20 to +50
15 - 90
+/- 3
(range 0-100)
36
15/40
(1)
1
COCl
2
Phosgene
1,00
-20 to +40
15 - 90
+/- 0.05
12
60/180
(2)
1
ETO
Ethylene Oxide
30.0
-20 to +50
15 - 90
+/- 1
36
50/240
(1)
36
H
2
Hydrogen
2,000
-20 to +50
15 - 90
+/-5 %
24
30/50
(1)
1
H
2
S
Hydrogen Sulfide
30.0
100
1,000
-40 to +50
15 - 90
+/- 1.5
(range 0-30)
36
15/30
(1)
1
HCl
Hydrogen Chloride
30.0
100
-20 to +40
15 - 95
+/- 0.4
(range 0-10)
24
30/150
(1)
24