10 Technical specifications
61
Type of gas
Measurement
range
(ppm)
XP Version
Temperature
range °C
% RH
Accuracy
(ppm)
Lyfe
(months)
Reponse
time
T
50
/ T
90
(s)
Storage
conditions
War
m-up
time
max
(h)
CO
2
carbon dioxide
5,000
5 %
10%
100%
-25 to +50
15 - 90
+/- 150
+/- 0.15%
+/- 0.3%
+/- 3%
60
15/30
(6)
2
R1234yf
(HFO-1234y f )
Tetrafluoropropene
2,000
0-100% LEL
-20 to +50
0 - 95
+/- 40 (from 0 to 50% FS)
+/- 100 (from 50 to 100% FS)
+/- 2% (from 0 to 50% LEL)
+/- 5% (from 50 to 100% LEL)
60
25/120
30/115
(5)
2
R134a
Tetraf luoroethane
2,000
-20 to +50
0 - 95
+/ - 40 (from 0 to 50% FS)
+/- 100 (from 50 to 100% FS)
60
40/170
(5)
2
R407f
2,000
-20 to +50
0 - 95
+/- 40 (from 0 to 50% FS)
+/- 100 (from 50 to 100% FS)
60
40/105
(5)
2
SF
6
Sulfur hexafluoride
2,000
-20 to +50
0 - 95
+/- 40 (from 0 to 50% FS)
+/- 100 (from 50 to 100% FS)
60
50/160
(5)
2
(5)
-40 85 °C
0 80 % RH
6 months maximum
(6)
4 20 °C
10 60 % RH
6 months maximum