EasyManua.ls Logo

Olympus BX-RFAA - Exciter Balancers U-EXBABG;EXBAUB;EXBAUG (for Exclusive Use with the BX-RFA;RFAA)

Olympus BX-RFAA
42 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
27
4
Exciter Balancer U-EXBABG/EXBAUB/EXBAUG (for exclusive use with the BX-RFA/RFAA)
}When a fluorescence image obtained by multiple excitation of U/B/G is observed with dual- or triple-band fluorescence
mirror units, use the exciter balancer to select the balance between the excitation light intensities of the fluorochromes.
* To use an exciter balancer with the BX-RFAA, remove the adjustment lever and attach it to the opposite end.
Clamping screw
Installation Procedure (Fig. 9)
1. Stand the adjustment lever @ of the exciter balancer vertically and insert
it the ND filter insertion slot with the same number as the slider on the
rear part of the right side of the illuminator.
· The insertion position is variable depending on the type of the exciter
balancer.
· Always insert the exciter balancer so that the clamping screw faces to-
ward the rear.
2. Tighten the clamping screw using the Allen screwdriver.
Operation Procedure
Observing a Double Stained Specimen
1. Set up normal reflected fluorescence observation.
2. Mount the fluorescence mirror units for double staining and engage them in the light path.
}Olympus standard products
Fluorescence Mirror Units
Exciter Balancer
Fluorescence Mirror
Units for Double Staining
Fluorescence Mirror
Units for Triple Mounting
U-EXBABG · U-DM-FI/TR2
· U-DM-FI/PI2
· U-DM-FI/TX2
· U-DM-DA/FI/TR2
· U-DM-DA/FI/PI2
· U-DM-DA/FI/TX2
U-EXBAUB · U-DM-DA/FI2
U-EXBAUG · U-DM-DA/TR2
· U-DM-DA/PI2
· U-DM-DA/TX2
# Due to its characteristics, the G-excitation
has a narrower intensity control range than
the U- and B-excitation. The intensity con-
trol range is also variable depending on
the status of specimen and variance in
mirror units’ characteristics.
# Lighting irregularities may be observed on
the upper and lower edges of the field dye
to the rotation angles of filters and the vari-
ance in mirror units’ characteristics. How-
ever, these lighting irregularities do not af-
fect the photographed area.
Fig. 9
Adjustment lever*
@
U-EXBAUG
U-EXBABG
U-EXBAUB

Related product manuals