EasyManua.ls Logo

Olympus BX51 - 3-5 Condenser

Olympus BX51
40 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
18
Fig. 29
3-5 Condenser
@
²
³
1 Centering the Condenser
(Figs. 29 & 30)
1. Turn the condenser height adjustment knob @ to raise the condenser to
its upper limit.
2. Focus on the specimen using the 10X objective.
#When using the U-SC3 swing-out condenser, move the top lens into
the light path.
3. Rotate the field iris diaphragm ring ² in the direction of the arrow so that
the diaphragm image comes inside the field of view.
4. Manipulate the condenser height adjustment knob @ to focus on the
diaphragm image.
5. Turn the two condenser centering screws ³ to move the iris diaphragm
image to the center of the field of view.
6. Gradually open the field iris diaphragm. The condenser is properly cen-
tered if the iris image is centered and inscribed in the field of view.
7. During actual use, open the field diaphragm slightly until its image cir-
cumscribes the field of view.
Effects of Field Iris Diaphragm (Fig. 30)
The field iris diaphragm restricts the diameter of the beam of light enter-
ing the objective and thus excludes extraneous light, improving image
contrast. The diameter of the field iris should e adjusted for objective
power to the extent that it just circumscribes the field of view. (See “Com-
patibility of Objectives and Condensers” on the next page.)
Fig. 30

Other manuals for Olympus BX51