EasyManua.ls Logo

Olympus BX51WI - User Manual

Olympus BX51WI
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
INSTRUCTIONS
BX51WI
FIXED-STAGE UPRIGHT
This instruction manual is for the Olympus Fixed-Stage Upright Microscope Model BX51WI. To
ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of
this microscope, we recommend that you study this manual thoroughly before operating the
microscope. Retain this instruction manual in an easily accessible place near the work desk for
future reference.
MICROSCOPE
A X 7 6 5 7

Table of Contents

Questions and Answers

  • M
    Mark KimAug 18, 2025
    What to do if the Olympus BX51WI bulb does not light?
    • J
      Jill ReedAug 18, 2025
      If the bulb does not light, it is likely that the bulb is burned out. Replace the bulb.
  • J
    Jack BrownAug 20, 2025
    What to do if Olympus BX51WI bulb lights intermittently?
    • T
      Teresa EllisAug 20, 2025
      If the bulb intermittently lights and goes out: * The bulb is nearly burnt out. Replace the bulb. * A cord or connector is not properly connected. Connect cords and plugs securely.
  • M
    Mary SmithAug 22, 2025
    How to clean dust from Olympus Microscope field of view?
    • C
      clintonmillerAug 22, 2025
      If dirt or dust is visible in the field of view and the cause is dirt/dust on the eyepiece, clean thoroughly.
  • D
    Dylan WilliamsAug 25, 2025
    How to fix dark field of view on Olympus BX51WI?
    • A
      Anthony ChristianAug 26, 2025
      If the field of view is dark when the bulb lights, there are several possible causes: * The aperture or field iris diaphragm is opened insufficiently. Open the aperture and field iris diaphragms. * The condenser is in too low a position. Adjust the condenser height. * The light path selector knob is set to the wrong position. Set the light path selector knob to the correct position. * The voltage selector knob is set to a low voltage position. Set it to the high voltage position.
  • P
    Pamela LopezAug 27, 2025
    Why is the Olympus Microscope image not sharp?
    • C
      catkinsAug 28, 2025
      If the visibility of the observed image is poor, the image is not sharp, the contrast is poor, or details are poorly visible, consider the following: * The objective in use is not designed for UIS2 (UIS) series. Replace it with a specified objective for UIS2 (UIS) series. * The revolving nosepiece is installed incorrectly. Secure it by pushing in the sliding dovetail all the way until the stopper. * The objective is engaged incorrectly in the light path. Make sure that revolving nosepiece clicks into place correctly. * There may be air in the objective front lens. Remove the air.
  • T
    thomasstantonAug 30, 2025
    What causes image glare on Olympus BX51WI?
    • J
      Jonathan MurrayAug 30, 2025
      Image glare can be caused by: * The condenser being set to too low a position. Adjust the condenser height. * The aperture iris diaphragm being closed too much. Open it sufficiently.
  • S
    Stacy WestSep 4, 2025
    Why does the Olympus BX51WI Microscope image appear to waver?
    • J
      Joshua MooreSep 4, 2025
      If the image appears to waver, it could be due to: * The revolving nosepiece being installed incorrectly. Secure it by pushing in the sliding dovetail all the way until the stopper. * The objective being engaged incorrectly in the light path. Make sure that revolving nosepiece clicks into place correctly. * The objective being placed incorrectly (may be loose) in the revolving nosepiece position. Insert the objective all the way into the revolving nosepiece position until it is stopped. * The condenser being centered incorrectly. Center it correctly.
  • J
    janice50Sep 7, 2025
    What causes one side of the Olympus BX51WI Microscope image to be blurred?
    • M
      Meghan RyanSep 7, 2025
      If one side of the image is blurred, check: * The revolving nosepiece is installed incorrectly. Secure it by pushing in the sliding dovetail all the way until the stopper. * The objective is engaged incorrectly in the light path. Make sure that revolving nosepiece clicks into place correctly. * The objective is placed incorrectly (may be loose) in the revolving nosepiece position. Insert the objective all the way into the revolving nosepiece position until it is stopped.
  • C
    Charles HolmesSep 9, 2025
    Why is brightness not changing on my Olympus BX51WI?
    • J
      James MarshSep 9, 2025
      If the brightness cannot be varied with the light intensity control: * No lamp bulb is installed. Attach a lamp bulb. * The lamp bulb is burnt out. Replace the lamp bulb. * The lamp housing output connector is unplugged. Plug the lamp housing output connector.
  • A
    andrewmcgrathSep 12, 2025
    Why is my Olympus Microscope field of view only slightly brighter?
    • W
      William DavisSep 12, 2025
      If the field of view becomes brighter only slightly although the voltage is increased: * The condenser is centered incorrectly. Center it correctly. * The condenser is in too low a position. Adjust the condenser height.

Summary

MODULE NOMENCLATURE

CONTROLS

Microscope Base and Power Supply

Operation and features of the microscope base and TH4 power supply unit.

Focusing Block Operation

Instructions for using the pre-focusing lever, fine adjustment knob, and frost switching lever.

Stage (IX-SVL2) Usage

Guide to placing specimens, moving the stage, and adjusting knob tension.

Revolving Nosepiece

Procedures for switching objectives using different revolving nosepiece models.

Observation Tube Adjustments

How to adjust interpupillary distance, diopter, and use eye shades.

Condenser Functions

Centering the condenser and using oblique illumination.

Immersion Objectives

Guidelines for using water immersion objectives and related caps.

TRANSMITTED LIGHT BRIGHTFIELD OBSERVATION PROCEDURE

OTHER OBSERVATION METHODS

Differential Interference Contrast (DIC)

Detailed steps for setting up and performing DIC observation.

Reflected Light Fluorescence

Instructions for reflected light fluorescence observation.

IR;DIC Observation

Observing specimens using infrared light and DIC.

Macro Reflected Light Fluorescence

Performing macro observation with reflected light fluorescence.

TROUBLESHOOTING GUIDE

Optical System Issues

Troubleshooting common problems related to the microscope's optical components.

Electrical System Issues

Resolving issues with bulbs, wiring, and power supply.

Adjustment Knobs, Tube & Stage Problems

Addressing problems with adjustment knobs, observation tube, and stage.

SPECIFICATIONS

OPTICAL CHARACTERISTICS

ASSEMBLY

Assembly Diagram

Diagram illustrating the sequence of module assembly for the microscope.

Detailed Assembly Procedures

Step-by-step instructions for assembling microscope components and modules.

LAMP HOUSING INSPECTION SHEET

PROPER SELECTION OF THE POWER SUPPLY CORD

Olympus BX51WI Specifications

General IconGeneral
TypeUpright Microscope
Optical SystemUIS2
CondenserAbbe condenser
IlluminationHalogen or LED illumination
Camera CompatibilityCompatible with various cameras
StageMechanical stage
FocusCoaxial coarse and fine focus knobs
NosepieceRevolving nosepiece
Observation TubesBinocular and trinocular observation tubes

Related product manuals