INSTRUCTIONS
BXFMA
MOTORIZED ILLUMINATION SYSTEM
WITH MOTORIZED FOCUSING UNIT
This instruction manual is for the Olympus Motorized Illumination System with Motorized Focusing
Unit, Model BXFMA. To ensure the safety, obtain optimum performance and to familiarize yourself
fully with the use of this system, we recommend that you study this manual thoroughly before
operating the system. Retain this instruction manual in an easily accessible place near the work
desk for future reference.
A X 7 6 1 5
BXFMA-F
U-FH
U-IFFH
U-BI30-2, U-TR30-2, etc.
U-LH100-3
Modules described in this manual

Questions and Answers

  • S
    Shawn MyersAug 16, 2025
    How to fix Olympus BXFMA Microscope when the bulb does not light?
    • M
      Matthew ZunigaAug 16, 2025
      If the Olympus Microscope bulb does not light, ensure the power cord of the BX-UCB (or TH4) is firmly plugged into the power outlet. Also, check that the main switch of the BX-UCB (or TH4) is set to ā€œIā€ (ON). Verify that the transmitted light illumination is correctly set, and if the bulb is burnt out, replace it with a new specified bulb.
  • B
    Brandon WalkerAug 19, 2025
    What to do if Olympus BXFMA visibility is poor and the image is not sharp?
    • A
      Angela AdamsAug 19, 2025
      If your Olympus Microscope visibility is poor, the image isn't sharp or the contrast is poor, ensure you're using the specified objective and eyepieces. Clean any dirt or dust from the specimen. Also, check that the aperture iris diaphragm is set properly.
  • F
    Felicia BoothAug 22, 2025
    How to fix an obscured or unevenly illuminated field of view on my Olympus Microscope?
    • D
      Daniel RiosAug 22, 2025
      If the field of view is obscured or not evenly illuminated on your Olympus Microscope, check the following: 1. Ensure the light path selector lever of the trinocular tube is correctly set according to the observation method. 2. Verify the revolving nosepiece is correctly mounted. 3. Center the aperture iris diaphragm. 4. Open the field iris diaphragm until it circumscribes the field. 5. Make sure the bulb is correctly mounted by pushing the pins of the halogen bulb all the way to the stop position.
  • D
    Darlene MonroeAug 26, 2025
    How to remove dirt or dust visible in the field of view of my Olympus Microscope?
    • C
      Carlos SolisAug 26, 2025
      If you see dirt or dust in the field of view of your Olympus Microscope, the dirt/dust may be on an eyepiece. Clean it thoroughly.
  • A
    Aaron TrevinoAug 28, 2025
    What to do if the specimen cannot be brought into focus on Olympus BXFMA?
    • A
      Alyssa JohnsonAug 28, 2025
      If the specimen cannot be brought into focus on your Olympus Microscope, adjust the hardware (or software) lower limit to an optimum height. If this does not solve the problem, re-place the specimen within the focusing range.
  • D
    Diana HolmesAug 31, 2025
    Why is the field of view dark on my Olympus BXFMA Microscope even when the bulb is on?
    • J
      jcervantesAug 31, 2025
      If the field of view remains dark even when the bulb operates on your Olympus Microscope, check if the aperture and field iris diaphragms are opened wide enough. Ensure the light path selector lever of the trinocular tube is in the correct position ( or ).
  • J
    Jay AndersonSep 3, 2025
    What causes the bulb to intermittently light and go out on an Olympus Microscope?
    • E
      Eric RodriguezSep 3, 2025
      If the bulb in your Olympus Microscope intermittently lights and goes out, it might be nearly burnt out, so replace it. Also, check all cords and connectors to ensure they are securely connected.
  • M
    Michael CareySep 6, 2025
    Why is one side of the image blurred on my Olympus BXFMA?
    • P
      Patricia BurkeSep 6, 2025
      If one side of the image appears blurred or wavers on your Olympus Microscope, ensure the specimen is correctly placed. Also, check that the revolving nosepiece is correctly mounted. If a heavier weight than restricted (3 kg) is applied to the revolving nosepiece mount area, reduce the number of modules to reduce the weight below the restriction.
  • J
    Jeffrey OwensSep 9, 2025
    Why does the Olympus Microscope bulb burn out so quickly?
    • A
      Abigail ReynoldsSep 9, 2025
      If the bulb burns out almost immediately in your Olympus Microscope, ensure you are using the correct bulb type.
  • G
    gwhiteSep 12, 2025
    Why doesn't the field of view match between both eyes on my Olympus Microscope?
    • S
      shafferalyssaSep 12, 2025
      If the field of view of one eye does not match that of the other on your Olympus Microscope, adjust the interpupillary distance and the diopter. Ensure that the eyepieces used on the left and right sides are the same type. When looking into eyepieces, do not stare at image from the beginning but see the overall field of view. It is sometimes recommended to turn your eyes away from the eyepieces, look far off and look into the eyepieces again.