EasyManua.ls Logo

Olympus BXFM-ILHS - User Manual

Olympus BXFM-ILHS
36 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
Loading...
INSTRUCTIONS
BXFM
FOCUSING UNIT
A X 7 6 1 3
This instruction manual is for the Olympus Focusing Unit model BXFM and associated modules.
To ensure safety, optimum performance and familiarize yourself fully with the use of the assembled
microscope, we recommend that you study this manual thoroughly before operating the unit. Re-
tain this instruction manual in an easily accessible place near the work desk for future reference.
This publication is printed on 100% recycled paper

Questions and Answers

  • D
    diazscottAug 13, 2025
    Why is the image blurry on my Olympus BXFM-ILHS?
    • M
      Mrs. Cheryl PenaAug 14, 2025
      Poor visibility, lack of sharpness, or poor contrast in the observed image can occur if the objective is not correctly engaged in the light path. Make sure that the revolving nosepiece clicks into place correctly. It could also happen if the objective or eyepiece in use are not the specified models. Replace them with the specified models.
  • K
    Kathleen MichaelAug 18, 2025
    What to do if the Olympus BXFM-ILHS Microscope bulb intermittently lights and goes out?
    • A
      Allison TurnerAug 18, 2025
      If the bulb intermittently lights and goes out, it might be near the end of its service life. Replace the bulb. Another potential cause is that a cord or connector is not properly connected. Ensure that all cords and plugs are securely connected.
  • H
    hjonesAug 21, 2025
    Why is the field of view obscured on my Olympus Microscope?
    • A
      Amanda DavidsonAug 21, 2025
      An obscured or unevenly illuminated field of view could be because the revolving nosepiece is not in a click position. Ensure it is set in a click position. Another cause could be that the light path selector knob of the trinocular tube is positioned halfway. Push the light path selector knob completely in or out. Also, check if the filter is not in a click position and set it correctly.
  • M
    Matthew DavisAug 25, 2025
    How to fix one side of the image being blurred on Olympus Microscope?
    • C
      Christopher GainesAug 25, 2025
      If one side of the image is blurred or looks like it's flowing, the specimen might be tilted. Correct the specimen positioning. Another reason could be that the objective is not correctly engaged in the light path. Ensure the revolving nosepiece clicks into place correctly. Also, if the light axis is tilted because the weight is heavier than specified, reduce the number of modules to or below the weight limit.
  • A
    Austin RoyAug 28, 2025
    Why does my Olympus Microscope LED light but the lamp bulb does not?
    • R
      Robert HowellAug 28, 2025
      If the LED lights but the lamp bulb does not, the bulb is likely burnt out. Replace the bulb. Another possible reason is that a cord or connector is not properly connected. Make sure to connect cords and plugs securely.
  • H
    hutchinsonnathanSep 1, 2025
    How to adjust Olympus BXFM-ILHS Microscope if field of view of one eye does not match that of the other?
    • C
      Christina MedinaSep 2, 2025
      If the field of view of one eye does not match that of the other, the interpupillary distance might be incorrect. Adjust the interpupillary distance. Alternatively, the diopter adjustment could be incorrect, so adjust the diopter. It is also possible that different eyepieces are used on the left and right. Change one eyepiece to match the other, ensuring both sides are the same type.
  • R
    Rhonda GreenSep 5, 2025
    Why is the coarse adjustment knob too heavy to rotate on my Olympus Microscope?
    • M
      Michael RichardsonSep 5, 2025
      If the coarse adjustment knob is too heavy to rotate, the rotation tension adjustment ring is likely too tight. Loosen the ring to an optimum tightness.
  • R
    Robert BellSep 9, 2025
    What to do if the focusing unit drifts down by itself on Olympus Microscope?
    • K
      Kevin ReidSep 9, 2025
      If the focusing unit drifts down by itself, causing a loss of focus during observation, the tension adjustment ring is likely too loose. Tighten the ring to an optimum tightness.

Summary

Important Safety Precautions

General Safety Guidelines

Critical safety guidelines for handling and operating the microscope system.

Maintenance and Storage

Using the Controls

Base and Power Supply

Details on voltage indication, power switch, and specimen placement.

Focusing Unit

Procedures for adjusting the focusing unit's coarse and fine adjustment knobs.

Observation Tube

Adjustments for interpupillary distance, diopter, and eye shades.

Reflected Light Brightfield Observation Procedure

Observation Methods (Using U-KMAS)

Reflected Light Nomarski DIC Observation

Procedure for setting analyzer and polarizer for Nomarski DIC observation.

Setting the DIC Slider

How to set the DIC slider for different objectives.

Reflected Light Simplified Polarized Light Observation

Procedure for simplified polarized light observation.

Troubleshooting Guide

Specifications

Optical Characteristics UIS2 (UIS) Series

Assembly

Proper Selection of the Power Supply Cord

Olympus BXFM-ILHS Specifications

General IconGeneral
Observation MethodBrightfield, Darkfield, Polarized Light
Magnification Range50x to 1000x
Camera PortTrinocular observation tube with camera port
FocusingManual coarse and fine focusing knobs
NosepieceManual, revolving nosepiece (typically 6 or 7 positions)

Related product manuals