Rotatable Clip Fixing Device
35
Chapter 3 Preparation, Inspection and Operation
Rotating the rotation grip also causes the
slider to rotate together. Therefore, be
sure to remove your finger from the slider
when rotating the rotation grip.
Rotate the rotation grip slowly. If it is
rotated quickly, the clip rotation may
jump.
If the endoscope is bent sharply, there
may be a delay between the rotation of
the rotation grip and that of the clip.
If the clip will not rotate smoothly,
straighten out the part of the insertion
tube extending from the forceps port as
straight as possible and rotate the
rotation grip (see Figure 3.12).
Figure 3.12
4. Press the clip against the targeted lesion.