Chapter 3 Cleaning, Disinfection and Sterilization Procedures
15
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
Reprocessing equipment parts and functions
For inspection of other equipment than that mentioned below, refer to the
instruction manual for the equipment being used.
Water-resistant cap (MH-553)
The water-resistant cap is attached to the electrical connector on the endoscope
to protect the connector from water penetration during reprocessing. For leakage
testing, the venting connector on the water-resistant cap must be connected to
the leakage tester (see Figure 3.2).
Figure 3.2
Channel plug (MH-944)
The channel plug is used to plug the openings of the instrument channel port,
air/water and suction cylinders during cleaning (see Figure 3.3).
Figure 3.3
Groove
Venting connector
Seal
Biopsy valve cap
Cylinder plug