EasyManua.ls Logo

Olympus EVIS LUCERA JF 260V - 3.1 Required reprocessing equipment

Olympus EVIS LUCERA JF 260V
80 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
14
Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3.1 Required reprocessing equipment
Preparation of the equipment
Prior to cleaning, disinfection or sterilization, prepare the equipment shown in
Figure 3.1.
Use basins which are at least 40 cm by 40 cm (16” by 16”) in
size and deep enough to allow the endoscope to be
completely immersed.
Figure 3.1
Detergent solution Large basins with tight fitting lids for
detergent and disinfectant solution
Personal protective
equipment
Clean water Large basins for rinsing and leakage
testing
Clean lint-free cloths
Sterile water
•30cm
3
(30 ml) syringe
Sterile lint-free cloths
70% ethyl or isopropyl alcohol
500 cm
3
(500 ml) containers
Sterile cotton swabs
Disinfectant solution Soft brush Small basins
•5cm
3
(5 ml) syringe
Channel-opening
cleaning brush
(MH-507)
Leakage tester
(MB-155)
Water-resistant cap
(MH-553)
Channel cleaning brush
(BW-20T)
AW channel cleaning
adapter (MH-948)
Suction cleaning
adapter (MH-856)
Channel plug
(MH-944)
Light source
(CLV-260/U40/U20)
Injection tube
(MH-946)
Maintenance unit
(MU-1)
Washing tube (MH-974)
Suction pump (KV-4/5)

Related product manuals