Chapter 3 Cleaning, Disinfection and Sterilization Procedures
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
3. Push the distal cover as far as possible toward the endoscope (see Figure
3.16).
4. Turn it counterclockwise until it stops (see Figure 3.16).
5. Pull the distal cover off the endoscope slowly and straightly (see Figure
3.16).
Figure 3.16
6. Reprocess the distal cover as described in Section 3.10, “Cleaning,
disinfection and sterilization procedures for distal cover”.