Chapter 3 Cleaning, Disinfection and Sterilization Procedures
55
EVIS LUCERA JF/TJF TYPE 260V REPROCESSING MANUAL
4. Sterilize the package according to the recommended ETO gas exposure
parameters described in Section 2.5, “ETO gas sterilization” and the
sterilizer manufacturer’s instructions.
5. Aerate the components following the minimum aeration parameters
specified in Section 2.5, “ETO gas sterilization”.
6. Store the components following the instructions given in Chapter 5, “Storage
and Disposal”.