SZ-CHI
@
C.
REPAIR
PROCEDURE
(3)
Set
the
jig.
@
Standard
stand
..
SZ-ST
0
Transformer
.....
TL2
CD
Eyepiece
........
G
series
eyepiece
®
Mirror
of
~20
or
more
(4)
Light
the
halogen
lamp
of
SZ-CHI.
(5)
Loosen
the
screws~
fixing
the
left
and
right·
polarizer@.
·
(6)
Adjust
focus
on
the
mirror
surface.
(7)
Turn
the
lefi
side
polarizer
while
monitoring
through
the
right
side
eyepiece,
and
tighten
t~e
screw
to
fix
th~
.Polarizer
at
the
position
where
the
view
bec9mes
most
dark.
(8)
Attach
the
quarter
wave
plate
to
the
SZ
microscopebottom
end,turn
it
gradually
while
monitoring
through
the
right
side
eyepiece,
and
stop
it
at
the
position
where
the
view
becomes
most
dark.
(9)
Turn
the
right
side
polarizer
while
monitoring·through
the
left
side
eyepiece,
and
tighten
the
screw
to
fix
the
polarizer
at
the
position
where
.the
view
becomes
most
dark.
(10)
Observe
the
mirror
surface
by
view-
ing
the
left
and
right
side
eyepieces
simultaneously.
Turn
the
quarter
wave
plate
and
check
if
the
cross
nicol
and
open
nicol
appear
simultaneously
in
the
left
and
right
visual
fields.
If
any
defect
occurs,
repeat
the
adjustment
from
(8)
C-3