SZ-CHI
®
C.
REPAIR
PROCEDURE
C-4
1-3
Adjusting
the
illumination
field
(1)
Remove
the
mirror
set
on
the
stage.
(2)
Insert
the
universal
standard
eyepiece
into
the
SZ
microscope
sleeve,
and
place
the
copy
of
the
chart
shown
on
the
left
on
the
stage
of
the
stand
as
a
specimen.
Universal
standard
eyepiece
(KN0048)
(3)
Adjust
focus
on
the
specimen
(so
that
the
focus
is
adjusted
in
all
the
zoom
range
to
ensure
exact
WD)
and
coincide
the
center
of
the
standard
eyepiece
with
the
center
of
the
specimen
by
moving
the
specimen.
(4)
Loosen
the
screws~
secur~ng
the
mirrors,
and
adjust
the
left
and
right
illumination
fields
by
turn-
ing
the
eccentric
pins~
so
that
the
concentric
circle
of
the
ilLumi-
nation
field
and
specimen
does
not
deviate.
(5)
Apply
an
adhesive
to
all
the
fixing
screws.
Adhesive
SHELLAC
(OT1131)
(6)
Mount
the
covers.