4-4 Differences in Operating Procedures of Software BX2-BSW
}The BX2-BSW software has been designed for controlling the motorized operations of the BX61 microscope system.
When the illuminator is controlled from the BX2-BSW, some operations may differ from the information shown in the
operation screens, tutorial and help of the software as described below.
1-2 Setting Procedure
1-2-2 Initial setting
2. Motorized unit setting (Device Type)
· Frame Enter “BX61F” or “BX61WI”. (Any can be entered without problem.)
· Reflected light illumination BX-RLAA
· Filter wheel U-FWR (The U-FWT cannot be used.)
· Autofocusing unit U-AFA1M
· Focus adjustment knob unit U-FH
(Note)
Text “enabled when the focus adjustment knob unit is installed” should be deleted.
3. Observation method setting (Microscopy Tab)
· Check only “EPI BF” (reflected light brightfield observation) and “EPI DF” (reflected light darkfield observation).
4. Mirror unit setting (Mirror Unit Tab)
· Enter “BF” or “DF”.
6. Turret optical device and top lens setting (Condenser Tab)
· Do not register anything.
7. Filter setting (Filter Tab)
· The U-FWT cannot be used.
1-2-4 Button function allocation
1. Button selection
Main Unit Tab
· Allocation to all of buttons 1 to 8 is possible with the PC screen operation.
· Even when button functions are allocated on the main unit, standalone operation on the main unit is impossible
(because the buttons are not present).
Other caution
· The microscope illustrations in the operation screens show the BX61.
BX2-BSW Help
Focus adjustment range setting
· The relationship between the objective moving direction and the Near/Far Limits is identical to that of the BX61WI.
· The height values entered as the Near and Far Limits cannot exceed the lower hardware limit (which is adjustable).