UES-40
Product Specification 1-4 ISSUE4
Saline system
2-1-5 Maximum values
Bipolar system
Cut: 90 W; Soft 1: 90 W; Soft 2: 90 W; Hard: L3
Monopolar system
Pure: 300 W; Blend: 250 W; Uro: 300 W; Coag 1: 120 W; Coag 2: 120 W;
Spray: 120 W
Saline system
Pure: 320 W; Blend: 320 W; Coag 1: 200 W; Coag 2: 80 W
2-1-6 Adjustment increments
Bipolar system
0 to 20 W: 1-W increments; 20 W to 30 W: 2-W increments; 30 W and over: 5-W incre-
ments
(Hard is adjustable in 1-level increments.)
Monopolar system
5-W increments
Saline system
5-W increments
2-1-7 Control switches
l Foot switch for UES-40 (MAJ-1258)
Controls the high-frequency output of monopolar and bipolar instruments (ON/OFF)
l Bipolar foot switch for UES-40 (MAJ-1259)
Controls the high-frequency output of bipolar instruments (ON/OFF)
l Handpiece output switch
Controls the high-frequency output of the handpiece (ON/OFF)
0 500 1000
1500
2000 2500
400
350
300
250
200
150
100
0
50
Pure
Blend
Coag 1
Coag 2
Saline output: Maximum setting
Load resistance ( )
Power (W)