UES-40
Troubleshooting 2-6 ISSUE4
(From previous
page)
Error code dis-
played
High-frequency current monitor dam-
age (Cut 01)
—132
Monitor board (J7) – Relay board (J2)
Used where the high-frequency cur-
rent is prone to leakage. (Touching
metal parts of patient’s bed or objects
such as external devices that have
fallen to the ground; used simulta-
neously with high-frequency grounded
electric scalpels [semi-floating], etc.)
Monitor board (J16) – Control board (J1)
High-frequency voltage-monitor dam-
age (Cut 02)
21435
Converter – Oscillation board (J1)
——
Converter – Oscillation board (J5)
Monitor board (J3) – Oscillation board (J7)
Monitor board (J6) – Oscillation board (J3)
Oscillation board (J4) – Control board (J5)
Monitor board (J7) – Relay board (J2)
Monitor board (J16) – Control board (J1)
Waveform-circuit damage (Cut 04) 1
Converter (Power-supply) 1 damage
(Cut 08)
3—12
Oscillation board (J4) – Control board (J5)
——
Converter (Power-supply) 2 damage
(Cut 10)
3—12
Oscillation board (J5) – Converter
——
Oscillation board (J4) – Control board (J5)
Converter (Power-supply) 3 damage
(Cut 20)
3—12
Oscillation board (J5) – Converter
——
Oscillation board (J4) – Control board (J5)
Patient-plate-circuit damage (Cut 40)
—132
Monitor board (J5) – Relay board (J1)
——
Monitor board (J16) – Control board (J1)
Problem
Parts That Might Have Caused the Problem
Oscillation
board
Monitor
board
Relay
board
Control
board
Panel
sheet
Converter Harnesses Other Other than main unit
*
1