Chapter 3 Installation and Connection
HIGH FLOW INSUFFLATION UNIT UHI-3
3.2 Connecting to an AC mains supply
Connect the UHI-3 main unit directly to a hospital grade AC
outlet (wall mains outlet).
Securely plug in the power cord so it will not be accidentally
dislodged during the operation.
Never apply excessive force to the power cord, such as
bending, straining, twisting or squeezing.
Do not use a 3-pin plug to 2-pin plug adapter as this will not
properly ground the UHI-3.
Always use the power cord provided with the UHI-3.
1. Confirm that the power is OFF.
2. Connect the power cord plug directly to a 3-pin hospital grade AC outlet
(wall mains outlet) which meets the power requirements indicated on the
electrical rating plate on the rear of the UHI-3 (see Figure 3.1).
Figure 3.1
Properly
grounded wall
mains outlet
Power plug