72
Appendix
HIGH FLOW INSUFFLATION UNIT UHI-3
UHI-3 main unit
Item Specification
Applicable gas •CO
2
gas for medical use.
Connection to gas cylinder via the Olympus
cylinder hose (PIN) for UHI-3 (MAJ-1080)
Connection to medical gas pipeline via the
Olympus medical gas pipeline adaptor for UHI-3
(MAJ-1084 (NIST), MAJ-1085 (DISS))
Pressure range of medical gas pipeline
–upper limit: according to ISO 7396 (1400 kPa)
–lower limit: 343.2 kPa (3.5 kgf/cm
2
)
Hose for medical gas pipeline is required to
comply with ISO 5359 (NIST or DISS)
Supply pressure indications Five steps by LEDs.
5: above 4.5 MPa
4: 3.5 to 4.4 MPa
3: 2.5 to 3.4 MPa
2: 1.5 to 2.4 MPa
1: 0.5 to 1.4 MPa
Alarm:
When the cylinder pressure is lower than
2.4 MPa, an intermittent alarm (1 second) will
sound.
Whenever the cylinder pressure drops 0.5 MPa,
an intermittent alarm (1 second) will sound.
When the cylinder pressure is 0 MPa, CO
2
insufflation will stop, the cylinder pressure
indicator will light red and an intermittent alarm
(3 second) will sound.