EasyManua.ls Logo

Olympus UHI-3 - Page 79

Olympus UHI-3
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Appendix
75
HIGH FLOW INSUFFLATION UNIT UHI-3
VOLUME
indicator
Display
method
0.0 – 19.9 L, 0.1 L step
20 – 999 L, 1 L step
Reset Pressing the reset button will return the display to
0.0 L.
Smoke evacuation function
(optional)
When the foot switch (optional) is depressed,
insufflation and suction are interlocked to
maintain abdominal pressure at the set pressure
while smoke evacuation is performed.
Inoperative when “LOW” is selected as
insufflation mode.
Inoperative when the abdominal pressure is
below 3 mmHg.
Classification
(electro-medic
al equipment)
Type of
protection
against
electric shock
Class I (3-pin power cord)
Degree of
protection
against
electrical
shock of
applied part
Type CF applied part:
Insufflation tube (MAJ-590, MH-397)
Suction tube (MAJ-591)
Weight 10 kg
Dimensions 295 mm (W) × 340 mm (D) × 150 mm (H) (housing
dimensions)
Power
requirements
Voltage (AC) 100 – 240 V
Frequency 50/60 Hz
Input 150 VA
Voltage
fluctuation
Within ± 10%
Suction pump Adjust to 400 to 300 mmHg.
Year of
manufacture
The year of manufacture is the second digit of the
serial number.
Item Specification
1212345

Table of Contents