VISERA ELITE II Quick Reference Guide
Disclaimer: Information provided in this guide is not exhaustive. Please refer to the complete Instructions For Use for each
item of equipment which should be thoroughlyread and understoodprior to handling, use and storage of the equipment.
TroubleshootingCamera HeadCamera Head
Adjust the image magnification and focus using either the
rings on the non-autoclavable version or buttons on the
autoclavable version.
There are three programmable buttons on the camera head.
To check what they are programmed to do press ‘info’ and
then scope information on the processor.
When each button is pressed its function will light up but the
screen needs to be closed before it will perform the function.
Non-autoclavable
Focus Ring
Zoom Ring
Autoclavable
Focus buttons
Zoom buttons
W - decreases
T - increases
No power
• Ensure everything is switched on.
• Check the isolation switch on the transformer
is turned on.
No image
• Ensure everything is switched on including the
monitor which can be switched off by accident.
• Unplug camera head, colour bars should appear.
If no colours appear, contact Olympus Service
Helpdesk.
Dark image
• Check condition of telescope.
• Check condition of light lead.
• Check fit of light lead is it firmly connected with
a click.
1
2
3
4
No. Nomenclature Description
1 User preset name Displays the user preset name.
2 Scope model Displays the model name of the camera head
3 Serial number Displays the serial number of the camera head
4 Remote switch Displays the function assigned to the remote switch