From the Setup Menu, press   to disregard changes to the current parameter value and 
exit the Setup Menu. 
  Presets: Save, Recall and Erase 
There are 20 presets in the instrument numbered 0 – 19.  The presets allow the user to save 
the parameters chosen during setup.  There are 3 preset actions.  The user can save, recall 
or erase.   
When a preset is saved, the current operating options are stored in one of the 20 selected 
presets.  The operating options include: 
•  Thermocouple Type 
•  Units 
•  Offset 
•  Open Lead Detection Status 
•  100% and 0% Span Settings 
•  Operating Mode: Fast Ramp, Slow Ramp, Step or Transfer 
To save a preset 
Press the 
(1.5s)
.  The preset number will start flashing.  Use the   to move to the 
preset number location you want to use to store the current operating options.  Press the 
.  The current operating options are now saved in the chosen preset location and the 
flashing stops. 
To recall a preset 
Press the   button.  “PRST” will begin to flash.  Use the  to move to the desired 
saved preset, 0-19.  When the desired preset is reached, press the preset button again to 
exit. The instrument will only display the numbers where presets are stored.  For example: 
If there are presets stored in 3 and 10 and all others are empty, in this case the   would 
only toggle between and display 3 and 10. 
 
To erase a preset 
To erase a preset it must first be recalled by following the “To recall a preset” steps above. 
Once the desired preset is recalled, press the 
(1.5s)
.  The preset number should now be 
flashing.  While the preset number is flashing, press 
(1.5s)
. “CLEAr” will appear on the LCD