4-18 
 
5.  Connect the Positive of the Source and Read channels to the positive input of the 
DMM.  Connect the negative of the Source and Read channels to the negative 
input of the DMM.  
6.  Press UUT   to turn the UUT on.  Disable the Auto-Power Off press 
(1.5s)
 
CAUTION 
Do not apply voltages greater than 83 mV DC to the UUT inputs. Voltages 
greater than 83 mV may damage the instrument.  
7.   
Insert the Straightened Paper Clip through 
the alignment access hole and gently press 
the calibration enable switch located on the 
circuit board to enter CAL mode. See Figure 
14 for location. 
Temporary calibration values are set to a 
gain of 1 and offset of 0 every time 
calibration is entered.  If the calibration is 
accepted and saved without entering new 
values, the temporary values are copied to 
the system values for use. 
Voltage Gain and Offset Alignment 
8.  The UUT display will indicate as follows: 
a.  Line 1: “-10.000” 
b.  Line 2: “CAL  1” 
The instrument is now sourcing  
-10.000 mV. 
9.  By using the   and/or 
 
 keys, 
adjust the instrument to match as close as 
possible the DMM display voltage. Use the 
 key to save the settings and advance to 
“CAL 2”.  Repeat this step to and including 
CAL 10. 
10.  CAL 11 and CAL 12 are used to set the Cold 
Junction Compensation, (CJC) of the 
Source, (Channel 1) and Read, (Channel 2).  These steps require the use of a 
customer supplied calibrated temperature measurement device. With the probe as 
close as reasonable to the channel 1 CJC, and temperature stabilized, enter the 
externally measured temperature in Celsius for channel 1.  Repeat this step for 
Channel 2, CAL 12. See figure 13 above.