Series Meters Quick Start Guide
6. Approvals Information
Power to Input/Output: 2300 Vac (3250 Vdc)
Power to Input/Output0F
: 1500 Vac (2120 Vdc)
Power to Relays/SSR Output: 2300 Vac (3250 Vdc)
Ethernet to Inputs: 1500 Vac (2120 Vdc)
Isolated RS232 to Inputs: 500 Vac (720 Vdc)
Isolated Analog to Inputs: 500 Vac (720 Vdc)
Analog/Pulse to Inputs: No Isolation
Measurement Category I
Category I includes measurements performed on circuits not directly connected to the Mains Supply
(power). Maximum Line-to-Neutral working voltage is 50Vac/dc. This unit should not be used in
Measurement Categories II, III, and IV.
Transients Overvoltage Surge (1.2 / 50uS pulse)
Input Power: 2500 V
Input Power1F
: 1500 V
Ethernet: 1500 V
Input/Output Signals: 500 V
ADDITIONAL INFORMATION:
FCC: This device complies with Part 15, Subpart B, Class B of the FCC rules, for option –EIP only.
RoHS II: The above product has been declared by the original supplier as Compliant. The manufacturer
of this item declares that the product complies with the EEE RoHS II Directive 2011/65/EC.26.
UL File Number: E209855
Low-voltage DC power option: Units configured for external low power DC voltage, 12–36Vdc.
Ibid.
Electrical Safety: 2014/35/EU (Low Voltage Directive) and Electrical Equipment (Safety) Regulations 2016
Safety requirements for electrical equipment for measurement, control, and laboratory
Double Insulation; Pollution Degree 2
Dielectric withstand Test per 1 min