Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 19 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
1.10 Check of internal connections & output impedance
In the following, the internal connections from the receivers to the 50 Ω output impedance are tested.
Additionally, the output of the Bode 100 will be checked.
Measurement Setup:
OPEN:
SHORT:
LOAD
(50 Ω):
Equipment Setup:
Figure 41: Internal connection check - measurement mode
Use the “One-Port” impedance measurement method and set the frequency to “Fixed” as well as
follows:
Figure 42: Internal connection check - measurement settings