Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 10 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
Figure 23: gain/phase accuracy - phase (°) measurement
1.5 Noise floor / Channel isolation
Measurement Setup:
Disconnect all BNC cables from Bode 100.
Do not connect anything to the front side BNC connectors!
Equipment Setup:
Please select the measurement type “transmission / reflection” and apply the following settings:
Figure 24: dynamic range / isolation - transmission / reflection measurement