Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 9 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
Figure 21: gain/phase accuracy - trace settings
Expected Measurement results:
Gain: 0 dB
Tolerance: max. ± 0.2 dB from 1 Hz to 50 MHz
Phase:
Tolerance: max ± 3° from 1 Hz to 50 MHz
If you measure a higher gain/phase deviation please send in the Bode 100 for calibration /
adjustment.
The following measurement results are an example and can be slightly different from the
measurement with other devices.
Figure 22: gain/ phase accuracy magnitude (dB) measurement

Other manuals for Omicron Lab Bode 100