1.3 Accuracy of output level
Measurement Setup:
Please connect the Source output of Bode 100 to a calibrated 50 Ω power meter.
Figure 10: output level accuracy - connection setup
Equipment Setup:
Please select the measurement type “Gain / Phase”, select the source settings as follows and start a
continuous measurement:
Frequency: 1 MHz – Fixed (or whatever frequency your power meter is made for)
Output level: 0 dBm / 13 dBm / -27 dBm
Figure 11: output level accuracy - measurement type setting
Figure 12: output level accuracy - measurement settings
Figure 13: output level accuracy - start continuous measurement