Bode 100 - Information
Detailed Functional & Calibration Check of Bode 100
Page 5 of 21
Smart Measurement Solutions
Smart Measurement Solutions
®
Equipment Setup:
Please select the measurement type “Gain / Phase”, select the source settings as follows and start a
continuous measurement:
Frequency: 10 MHz - Fixed
Output level: 0 dBm
Figure 7: output frequency accuracy - measurement type setting
Figure 8: output frequency accuracy - measurement setting
Figure 9: output frequency accuracy - start continuous measurement
Expected Measurement result:
Measured frequency: 10 MHz
Maximum tolerance for Bode 100 Revision 1:
± 15 ppm ± 150 Hz @ 10 MHz (< 1 year after adjustment)
± 25 ppm ± 250 Hz @ 10 MHz (< 3 years after adjustment)
Maximum tolerance for Bode 100 Revision 2:
± 2 ppm ± 20 Hz @ 10 MHz (< 1 year after adjustment)
± 4 ppm ± 40 Hz @ 10 MHz (< 3 years after adjustment)
If you measure a higher frequency deviation, please send the Bode 100 in for calibration / adjustment.