+ Draw air into the fill syringe equal to the
amount of insulin indicated in the individual’s
Diabetes Management Plan.
+ Depress air into the vial of insulin.
+ Turn the vial and syringe upside down.
+ Withdraw insulin from the vial and fill the
syringe with the amount of insulin indicated in
the individual’s Diabetes Management Plan; fill
at least to the MIN line.
+ Remove any air bubbles from the syringe.
+ Insert the needle straight down into the fill port
on the underside of the Pod. To ensure proper
fill, do not insert fill syringe at an angle into the
fill port.
+ Completely empty the syringe into the Pod.
+ The Pod will beep twice, indicating that the
System is ready to proceed.
WARNING: NEVER inject air into the fill port.
Doing so may result in unintended or interrupted insulin delivery.
1.3
1.4
WARNING: NEVER use a Pod if you hear a crackling noise or
feel resistance when you depress the plunger. These conditions
can result in underdelivery of insulin.
HOW TO CHANGE THE POD