7
• Ensure that this monitor has acclimated to room temperature before taking
a measurement. Taking a measurement after an extreme temperature
change could lead to an inaccurate reading. It is recommended that you
wait for approximately 2 hours for the monitor to warm up or cool down
when the monitor is used in an environment within the temperature
specified as operating conditions after it is stored either at the maximum or
at the minimum storage temperature. For additional information on
operating and storage/ transport temperature, refer to section 9.
• DO NOT use this monitor after the durable period has ended. Refer to
section 9.
• DO NOT crease the arm cuff or the air tube excessively.
• DO NOT fold or kink the air tube while taking a measurement. This may
cause an injury by interrupting blood flow.
• When the monitor malfunctions, it may get hot. DO NOT touch the monitor if
it occurs.
• To unplug the air plug, pull on the plastic air plug at the base of the tube, not
the tube itself.
• ONLY use the AC adapter, arm cuff, batteries and accessories specified for
this monitor. Use of unsupported AC adapters, arm cuffs and batteries may
damage and/or may be hazardous to this monitor.
• ONLY use the approved arm cuff for this monitor. Use of other arm cuffs
may result in incorrect readings.
• Inflating to a higher pressure than necessary may result in bruising of the
arm where the cuff is applied. NOTE: refer to “If your systolic pressure is
more than 210 mmHg” in section 3 for additional information.
AC Adapter (optional accessory) Handling and Usage
• Fully insert the AC adapter into the outlet.
• When unplugging the AC adapter from the outlet, be sure to safely pull from
the AC adapter. DO NOT pull from the AC adapter cable.
• When handling the AC adapter cable:
Do not damage it. / Do not break it. / Do not tamper with it.
DO NOT pinch it. / Do not forcibly bend or pull it. / Do not twist it.
DO NOT use it if it is gathered in a bundle.
DO NOT place it under heavy objects.
• Wipe any dust off of the AC adapter.
• Unplug the AC adapter when not in use.
• Unplug the AC adapter before wiping this monitor.