GUIDANCE AND MANUFACTURERS DECLARATION
Recommended separation distance between portable and mobile RF communications
equipment and the OMRON BP742N including AC-adapter
OMRON BP742N including AC-adapter is intended for use in an electromagnetic environment in
which radiated RF disturbances are controlled. The customer or the user of this OMRON BP742N
including AC-adapter can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the
OMRON BP742N including AC-adapter as recommended below, according to the maximum
output power of the communications equipment.
Output Power of Transmitter
in Watt
Separation distance according to frequency of
transmitte
r in meter
150 kHz to 80
MHz
d = 1.2 P
80 MHz to 800 MHz
d = 1.2 P
800 MHz to 2.5GHz
d = 2.3
P
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts (W) according
to the transmitter manufacturer.
Note: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
Note: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorpt
ion and reflection from structures, objects, and people.
37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Omron BP742N and is the answer not in the manual?

Omron BP742N Specifications

General IconGeneral
TypeAutomatic
Cuff size23 - 43 cm
Product colorWhite
Memory functionYes
Number of users2 user(s)
Memory registers100
Placement supportedUpper arm
Mean arterial pressureNo
Quantity per pack1 pc(s)
Display typeAnalog