Weight Main unit: Approx. 0.52 kg
(not including accessories and options)
AC adapter: Approx. 120 g
Batter
: Approx. 0.1 k
Displa
7 se
ment LCD
Protection Class
Class II (AC Adapter)
Internal powered equipment (when operating with
batter
onl
De
ree of Protection T
pe BF
Applied part
: Cuff
Operation Mode Continuous operation
MDD Classification Class II a
Packaging contents Main unit, 2 Arm cuffs (M size/L size), AC Adapter,
Batter
pack, Instruction Manual, Guarantee card
Power supply
AC adapter Input voltage range: AC 100 V to 240 V
Frequency: 50/60 Hz
Output voltage range: DC 6 V ±5%
Rated output current: 1.6
Rechargeable battery Type: 3.6 V, 1900 mAh
Number of operation cycles when fully charged: 300
Measurement conditions
• New battery fully charged
• Ambient temperature of 23°C (73.4°F)
• Using M-size cuff
• SYS 120 / DIA 80 / PR 60 (Inflation setting: AUTO)
• One 5-minute cycle consisting of “cuff measurement
time + wait time”
Environmental Conditions
Operating Conditions Temperature range: 5 to 40°C (41 to 104°F)
Humidity range: 15 to 85%RH (not condensed)
tmospheric pressure: 700 to 1060hPa
Storage and transportation Temperature range: -20 to 60°C (-4 to 140°F)
Humidity range: 10 to 95%RH (not condensed)
tmospheric pressure: 500 to 1060hPa
Non-Invasive Blood Pressure (NIBP)
Measurement technolo
Oscillometric
Measurement method D
namic Linear Deflation method
Pressure displa
ran
e 0 to 300 mmH
Pressure displa
accurac
Within ±3 mmH
NIBP measurement range SYS 60 to 250 mmHg
DIA 40 to 200 mmHg
Pulse rate 40 to 200 /min
NIBP accuracy* Maximum mean error within ±5 mmHg
Maximum standard deviation within 8 mmH