Storage and Handling
Please review the following storage and handling instructions:
Store the control solution at 2-30 °C (36-86 °F).
Do not refrigerate or freeze.
If the control solution is cold, do not use until it has warmed to room
temperature.
Use before the unopened expiration date that is shown on the bottle.
Note: The expiration date is printed in Year-Month format. 2014-01 means
January, 2014.
Each bottle of control solution can be used for 6 months after you first open it.
Record the opened and the resulting expiration date on the bottle label.
Control Solution Precautions
For in vitro diagnostic use. The control solution is for testing only outside the
body. Do not swallow or inject.
Shake well before using.
Control solution tests are designed to be accurate only when tested between
10 and 40 °C (50-104 °F).
The control ranges shown on the test strip vial (or on the foil pouch) are not
recommended ranges for your blood glucose level. Your personal blood
glucose target ranges should be determined by your diabetes healthcare
professional.
Do not touch the test strip with the tip of the control solution bottle.
Use only the same brand of control solution that was provided with your kit.
See the control solution insert for more details.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the On Call Extra and is the answer not in the manual?

On Call Extra Specifications

General IconGeneral
Sample Size0.4 µL
Test Time4 seconds
Blood SampleCapillary whole blood
Battery2 x AAA batteries
DisplayLCD
Battery LifeApproximately 1, 000 tests
Automatic Power Off2 minutes
Operating Humidity10-90% RH
Measurement Range20 - 600 mg/dL (1.1 - 33.3 mmol/L)
Memory450 tests
Operating Temperature5°C to 45°C (41°F to 113°F)