System Accuracy Results for Glucose
Concentrations between 1.8 mmol/L and
27.3 mmol/L
Percent (and number) of meter results that match the
laboratory test
Within ±0.83 mmol/L or ±15%
(99.0%) 594/600
NOTE: Where 1.8 mmol/L represents the lowest glucose
reference value and 27.3 mmol/L represents the highest
glucose reference value (YSI value).
Regression Statistics
Samples were tested in duplicate on each of three test
strip lots. Results indicate that the OneTouch®Verio®
System compares well with a laboratory method.
# of Subjects # of Tests Slope Intercept
(mmol/L)
100 600 0.99 0.29
95% CI
Slope
95% CI Intercept
(mmol/L)
Std. Error (S
y.x
)
(mmol/L)
R
2
0.98 to 0.99 0.20 to 0.37 0.55 0.99