Detailed information
about your system
8
121
Precision
Within Run Precision (300 Venous Blood Tests per Glucose
Level)
Target
Glucose
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
2.2 2.86 0.07 2.44
5.6 6.03 0.11 1.76
7.2 8.09 0.16 2.00
11.1 11.49 0.24 2.08
19.4 21.22 0.43 2.01
Results show that the greatest variability observed
between test strips when tested with blood is 2.44% or
less.
Total Precision (200 Control Solution Tests per Glucose
Level)
Glucose
Level Ranges
(mmol/L)
Mean
Glucose
(mmol/L)
Standard
Deviation
(mmol/L)
Coecient
of Variation
(%)
Low (2.11–3.44) 2.19 0.05 2.08
Mid (5.67–7.67) 6.54 0.12 1.88
High (16.56–22.39) 19.01 0.36 1.91