Ref1109 User Manual • • • 15
The hoist needs to be checked once a year according to the local-guidelines (UK LOLER every 6 months).
This needs to be done by OpeMed or another OpeMed authorized institution. OpeMed offers you the
possibility of an annual inspection by means of a Service Agreement.
If maintained by an authorised person the OT200 hoist should have a sticker tells you when the hoist is
due for inspection. For more information about inspections you can always contact OpeMed.
7.4 If the OT200 Does Not Function
• Check whether the emergency cord is pulled out
• Check whether the batteries are sufficiently charged.
• Check whether the plug of the hand-held control is connected correctly.
• Check whether the plug of the hanger bar control is connected correctly.
• Check whether the belt is slack, if necessary tighten it.
• Check whether the belt has been pulled up too much, the hoist will only go down.
• Check whether the belt has been rolled out too much, the hoist will only go up.
If the hoist is still not working, alarm the check technical service department.
7.5 Disposal
For disposal the OT200 can be returned to OpeMed. This is also possible with batteries, chargers, slings
etc. If you choose to dispose the batteries yourself, be aware of the fact that these batteries are chemical
waste.
8 OT200 Specification
8.1 Technical Summary
Maximum hoist capacity 200kg (400kg for OT400)
Sound pressure level 47 dB (A)

Other manuals for OpeMed OT200