3
Hold the calibrator with the
LOT number (glucose) or
calibration CODE
(ß-Ketone) facing you.
4
Insert the calibrator into
the electrode port. Push it in
until it stops.
The Display Check shows on
the display window,
followed by the time,
month, and day.
Next, the LOT number
(glucose) or calibration
CODE (ß-Ketone) shows on
the display window.
a. Example
of glucose LOT
number: LOT 13758
b. Example
of ß-Ketone
calibration CODE:
CODE JC7
Then you hear the beeper, if
the beeper is ON.
Glucose
ß-Ketone
Calibrating
Your Monitor
25
133-221 Manual,G3b,Aus 3/17/05 11:15 AM Page 28

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Optium Xceed and is the answer not in the manual?

Optium Xceed Specifications

General IconGeneral
Sample Size0.6 microliters
Test Time5 seconds
Battery LifeApproximately 1000 tests
Blood SampleCapillary whole blood
Memory Capacity450 results
Operating Temperature10°C to 40°C
Operating Humidity10% to 90% RH
Power Source2 x AAA batteries