Understanding
Your Result
Compare the control result to
the "Expected Results for Use
with MediSense Control
Solutions" range printed on:
The blood glucose electrode
instructions for use,
for glucose control
solution testing.
The blood ß-Ketone electrode
instructions for use and foil
packet, for ß-Ketone control
solution testing.
If the control result falls
within the range:
What It Means:
Your monitor and electrodes
are working correctly.
If the control result does
not fall within the range:
What It Means:
Your monitor and electrodes
may not be working correctly.
What to Do:
Repeat the test with a new
electrode and ensure the control
solution testing instructions are
followed completely. If the result
is still not within the printed
range, contact Customer Service.
Control
Solution Testing
64
133-221 Manual,G3b,Aus 3/17/05 11:15 AM Page 67

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Optium Xceed and is the answer not in the manual?

Optium Xceed Specifications

General IconGeneral
Sample Size0.6 microliters
Test Time5 seconds
Battery LifeApproximately 1000 tests
Blood SampleCapillary whole blood
Memory Capacity450 results
Operating Temperature10°C to 40°C
Operating Humidity10% to 90% RH
Power Source2 x AAA batteries