50
You hear the beeper, if the
beeper is ON.
The display window shows
the status bar.
Then the display window
shows the countdown.
Note: Do not remove the
electrode from the monitor
or disturb it during
the countdown.
Important: If the countdown
does not star
t:
What It Means:
You might not have applied
enough blood to the electrode.
What to Do:
Apply a second drop of blood
to the electrode within 30
seconds of the first drop.
If the countdown still does not
start or if more than 30
seconds have passed, discard
the electrode, turn off your
monitor, and try again with a
new electrode.
Monitoring
Blood ß-Ketone
Status bar
Countdown