21 Error messages
21.1 Internal error messages
Display: Lxx
Internal errors arise when recording and processing measured values in the entire
measurement route from the sensor to storing in memory or on the display.
00 Measured value is ok
01 AD conversion faulty
0
2 Communication error
03 Over/underflow
05 Wire break
06 Sensor type dependent
0
7 Expansion card not available
21.2 External error messages
Display: Sxx
External errors occur in "intelligent sensors" and are sent to the OTT netDL.
21.3 Sensor error messages
Analog input
Internal errors: 01 Analog-digital conversion faulty
03 over-/underflow
0 1,25 V: U
in
< –0,1 V or U
in
> 1,375 V
0 10 V: U
in
< –0,5 V or U
in
> 11,0 V
0 5 V: U
in
< –0,5 V or U
in
> 5,5 V
0 50 mV: U
in
< –5 mV or U
in
> 55 mV
0 20 mA: I
in
< –2 mA or I
in
> 22 mA
4 20 mA: I
in
< +2 mA or I
in
> 22 mA
Pt 100: t < –100 °C or t > 140 °C
Potentiometer: R < –10 % or R > 130 %
04 Timeout during measurement
05 Wire break
External errors: None
Pulse input
Internal errors: None
External errors: None
Kalesto
Internal errors: 01 Set minimum signal level underrun
02 Communication error
05 Wire break
External errors: 01 Reflexion level too low
02 Multiple-target situation. No measured value can
be determined
04 Measurement range exceeded, measured
value < 1.50 m or > 30.00 m
08 A constant measured value cannot be determined
Sonicflow
Internal errors: 01 over-/underflow
02 Communication error
04 Wire break
05 Missing water level
External errors: 01 Signal level underrun
51