English; Version 08 | 2019-04-24
remOVE DC Impulse 400.01
Instructions for use
19 / 32
14.1 Wipe disinfection
For cleaning surfaces of the device, please use approved cleaning/disinfection sup-
plies and only in accordance with instructions by the respective manufacturer. Ob-
serve specifications regarding concentration, temperature and exposure time.
Apply alcohol-based cleaning/disinfection agent. Do not use benzyl-alcohol-based
and/or any other agents because it can cause damages to the remOVE DC Impulse
materials.
14.2 Instructions
1. Prepare the cleaning/disinfectant agent per the manufacturer’s guidelines.
2. Wipe down the equipment using a cloth with surface disinfectant. Clean gross
contamination first and then uniformly treat all the surfaces.
3. Wipe the surfaces ensuring that they are uniformly treated. Comply with the
action time of the disinfectant specified by the manufacturer.
4. Wet a sponge or cloth in clean water and wipe off the cleaning/disinfection
agent.
5. Dry the device using a clean, lint-free cloth.
6. Check by visual inspection all surfaces of the equipment. If soil remains visible,
repeat the entire cleaning/disinfection procedure.
14.3 Safety Instructions
Before cleaning the remOVE DC Impulse, disconnect the device from the power
supply.
Do not use flammable or explosive cleaning or disinfecting solutions. Make sure no
fluids enter the device casing. Non-compliance may result in burning hazard and/or
electrocution.
Only clean and disinfect the product manually. Do not sterilize the product under
any circumstances.