remOVE DC Impulse 400.01
Instructions for use
25 / 32
Compatibility
remOVE DC Cutter Set 12 / 14
Art.-Nr. 400.02.01 / 400.02.02
Foot-activated switch
Art.-Nr. 400.03
DC cord
Art.-Nr. 400.04
Power cord
Art.-Nr. 400.10.XX
Environmental condi-
tions for operation,
transport and storage
Operation
Storage
Transportation
Temperature
+10 °C to +40 °C
0 °C to +50 °C
-20 °C to +50 °C
Relative humidity
25 % to 75 % RH,
non-condensing
0 % to 90 % RH,
non-condensing
0 % to 90 % RH,
non-condensing
Air pressure
700 to 1060 hPa
500 to 1060 hPa
500 to 1060 hPa
Operation level
4000 m
-
-
20 Electromagnetic compatibility (EMC)
Make sure no electronic devices which could be compromised by electromagnetic
fields are present in the vicinity of the remOVE DC Impulse. A compromised device
could lead to malfunction and/or failure of respective device and thus may endan-
ger user and/or patient.
Medical electrical equipment needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the accompanying documents.
Guidance and manufacturer’s declaration
Electromagnetic emissions
The remOVE DC Impulse is intended for use in the electromagnetic environment specified be-
low. The customer or the user of the remOVE DC Impulse should assure that it is used in such
an environment.
Measurements of electromagnetic emissions
Compliance
RF emissions CISPR11
Group 1
RF emissions CISPR11
Class A
Harmonic emissions IEC 61000-3-2
not applicable
Voltage fluctuations / flicker emissions IEC 61000-3-3
complies