Instructions for use
remOVE DC Impulse 400.01
English; Version 08 | 2019-04-24
Guidelines and manufacturer's declaration
Electromagnetic immunity
The remOVE DC Impulse is designed for use in an electromagnetic environment as specified be-
low. The user must ensure that the device is operated in such an environment.
Compatibility test
IEC 60601 test level
Compliance level
Electrostatic discharge (ESD) ac-
cording to IEC 61000-4-2
± 8 kV contact discharge
± 8 kV contact discharge
± 2,4,8 and 15 kV air
discharge
± 2,4 and 8 kV air discharge
Fast transient elec. disturbance
values/bursts according to IEC
61000-4-4
± 2 kV for power cables
± 2 kV power cables
Surges according to 61000-4-5
± 0.5 kV, ± 1 kV line to line
voltage
± 0.5 kV, ± 1 kV line to line
voltage
± 0.5 kV, ± 1 kV, ± 2 kV
line to earth voltage
± 0.5 kV, ± 1 kV, ± 2 kV line
to earth voltage
Short interruptions according to
IEC 61000-4-11
0 % U
T
; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 31
0 % UT; 0.5 cycle
At 0°, 45°, 90°, 135°, 180°,
225°, 270° and 31
0 % U
T
; 1 cycle and 70 %
U
T
; 25/30 cycles
single phase: at 0°
0 % U
T
; 1 cycle and 70 %
U
T
; 25/30 cycles
single phase: at 0°
Voltage dips according to IEC
61000-4-11
0 % U
T
; 250/300 cycle
0 % U
T
; 250/300 cycle
Power frequency magnetic field
(50/60 Hz) according to IEC
61000-4-8
30 A/m
30 A/m
Please note: U
T
is the mains voltage prior to applying the test level.