13 Error message
The remOVE DC Impulse is equipped with a failure detection system. Technical
failures are indicated via the status display. Failure detection is implemented to
ensure user/patient safety.
If a failure message occurs, the device casing may not be opened. Life-threatening
electrocution could occur.
If a failure message occurs, the device may no longer be used in order to ensure
the safety of users and/or patients. If a failure occurs, all eight LEDs of the status
display are flashing orange, see figure below.
Figure 12: Status display flashes orange during failure message.
If a failure message is displayed by the remOVE DC Impulse, acoustic contact de-
tection between instrument and clip is deactivated. A DC pulse can no longer be
generated, and the internal energy storage of the remOVE DC Impulse can no
longer be recharged.
In the case of failure message or malfunctioning of the device, please contact man-
ufacturer.
14 Cleaning and disinfection
As part of your responsibility to ensure hygiene and cleanliness of all product com-
ponents during use, please make sure that only suitable devices and procedures
validated specifically for this product are used for cleaning and disinfection of the
product. Please follow the respective local and/or national legislation as well as
hygiene regulations pertaining to medical practices or hospitals.